Remove the protective cap from the medical container and disinfect the cap.
Remove the protective cap from the spike.
Insert the spike.
Open the plug cover.
Turn on the syringe and vacuum the desired volume.
Important: The solution from a full syringe to a container should only be injected with the correct side up. Risk of excessive pressure and poor air leak.
Materials used
Structure: ABS/SAN;
Occlusion cover: PP
Protective cap and valve seal: PE
Filter: acrylic copolymer in nylon body material
Valve disc: silicone
Application duration
Consult the instructions for use of the medicine, as well as the hygiene and safety regulations.
Notices
The reuse of single-use devices creates a potential risk for the patient, or the user. It may lead to contamination and/or insufficient functional capacity. Contamination and/or impairment of the device's functionality may lead to injury, illness or death of the patient.
Filtration - Due to a possible level of absorption >15%, do not use spikes with liquid/particle filters (ventilation filter is not essential) with: corticotropin, cosintropine, cyanocobalamin, ergonovines, malleates, glycagine, insulin, vasopressin.
Substances that cause stress cracks: Several organic and inorganic substances with surface activity, solvents (e.g. macrogol/polyethyleneglycol and/or benzyl alcohol) in drugs or active substances can cause stress cracks or strain in plastics.